Publication

Use of a VBX balloon expandable stent-graft for management of right brachiocephalic vein stenosis.

Jones, Robert G
Willis, Andrew P
Abstract
Here we describe the use of a VBX balloon expandable stent-graft (WL Gore, Flagstaff, AZ, USA) to treat a right brachiocephalic vein stenosis in a haemodialysis patient prior to ipsilateral upper limb arterio-venous (AV) fistula formation. Balloon expandable stent-grafts are unsuitable for treating peripheral fistula stenoses due to their susceptibility of being crushed. The right brachiocephalic vein is both relatively short in comparison to the left and is less susceptible to extrinsic compression and the use of such a device to treat stenosis here allows for very accurate placement and restoration of luminal diameter. The advantages and disadvantages of using these devices in haemodialysis access circuits are also discussed herein, in what we believe to be the first report of the application of a dedicated commercially available balloon expandable stent graft in a haemodialysis patient.
Affiliations
Citations
Altmetric:
Date
2021-03-05
Type
Article
Subject
Collections
Citation
Jones RG, Willis AP. Use of a VBX balloon expandable stent-graft for management of right brachiocephalic vein stenosis. J Vasc Access. 2022 May;23(3):477-480. doi: 10.1177/1129729821998835. Epub 2021 Mar 5
Journal / Source Title
Publisher’s statement
Note / Copyright